All parameters described on the next tables are DWord data types and their bits can be accessed separately.

ALARM1

Bit

Description

0

No alarm

1

Calibration data corrupt

2

Battery alarm

3

Therapy data corrupt

4

Therapy data and pump settings corrupt

5

Battery voltage too low

6

Battery not in pump

7

Battery flat at power on

8

Battery cover open

9

Syringe empty

10

Pressure too high

11

Syringe drive blocked

12

Standby time expired

13

Volume infused

14

Infusion time expired

15

Claw malfunction

16

Pressure sensor defect

17

Syringe holder open

18

Syringe not inserted correct

19

KVO end

20

Pressure alarm upstream sensor

21

Dummy

22

Drop alarm, summary of all drop alarms

23

Drop alarm: no drops

24

Drop alarm: too less drops

25

Drop alarm: too many drops

26

Drop alarm: Free flow

27

Drop alarm: no drop sensor con. but required

28

Air alarm, summary of all air alarms

29

Air alarm: Air bubble too large > 0,3 ml

30

Air alarm: Air rate exceeded > 4 ml/h

31

Air alarm: sensor test faulty

 

ALARM2

Bit

Description

0

Dummy

1

Data-lock unauthorized manipulation

2

X hour limit is reached, only for PCA operation

3

SGC-End alarm

 

STATUS1

Bit

Description

0

1: Pump switched on

1

1: Mains power operation

2

1: Standby active

3

1: Start-up menu at pump active

4

1: Prime active

5

1: Drive in parking position

6

1: Pump is ready to start

7

1: Pump is running or infusing

8

1: Therapy has started

9

1: Ready to start bolus

10

1: Manual bolus is active

11

1: Volume bolus is active

12

1: KVO function released

13

1: KVO active

14

1: Data lock active

15

1: Pump is operating in dose mode

16

1: Dose bolus with body weight

17

1: Patient data stored in pump

18

1: Manual operation at pump

19

1: act. Alarm is quit

20

1: act. pre alarm is quit

21

1: Space-Control is connec. to pump

22

1: Air sensor is switched off

23

0: Piggyback active or 1: primary active

24

1: Editor at pump is active, that is, rate is edited

25

1: intermittent bolus is active

26

1: minimum one alarm is active

 

PREALARM

Bit

Description

0

No pre alarm

1

Disposable not inserted

2

Volume to be delivered near end

3

Time near end

4

Battery near flat

5

KVO operation

6

Data lock unauthorized manipulation

7

Incompatible CAN-BUS device detected

8

Piggyback: secondary volume infused

9

SGC pre alarm